BC Biomedical SA-2001 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Mesure, test et contrôle BC Biomedical SA-2001. BC Biomedical SA-2001 User manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - USER MANUAL

ELECTRICAL SAFETY ANALYZER SA-2001 SA-2001-INTL SA-2001-AUS USER MANUAL

Page 2

2 AVAILABLE MODELS: SA-2001 – STANDARD MODEL WITH HOSPITAL-GRADE NEMA 5-15P LINE PLUG AND HOSPITAL-GRADE NEMA 5-20R DUT TEST RECEPTACLE FOR USE I

Page 3 - BC BIOMEDICAL

3 BC20-20200 REPLACEMENT SA-2001-INTL DUT INTERNATIONAL RECEPTACLE ADAPTER BC20-20221 SCHUKO GROUNDING ADAPTER BC20-41337 RS-232

Page 4 - WARNING - LIQUIDS

4 This section looks at the layouts of the SA-2001 Models and gives descriptions of the elements that are present. For all models, the user contro

Page 5 - CAUTION - FUSE

5 SA-2001-INTL Layout SA-2001-AUS Layout (SA-2001-INTL-R shown) The

Page 6 - NOTICE – SYMBOLS

6 Display The main information in the system is presented in the 3-1/2 digit LCD display. This data is provided as simple meter readings with the m

Page 7 - NOTICE – CONTACT INFORMATION

7 Connectors There are two connectors for test cables on the unit. One is for the Chassis lead and the other is for one of two different leads used

Page 8

8 Communications (Optional) On Analyzer models with the –R option, a DB-9 RS-232 Communications p

Page 9

9 The SA-2001 Models allow the user a great deal of flexibility in testing. Any of the basic tests can be run and in almost any sequence. The infor

Page 10

10 Device Current With the Device Current function selected, the display will show the current draw of the DUT. The Receptacle should be configured

Page 11

11 External Point to Point Measurements The unit has the ability to measure Leakage Current and Resistance between two points, utilizing

Page 13

12 The communication protocol provides a means to completely configure and use the Analyzer from a PC or other device with RS-232 communications int

Page 14

13 case characters in the command summary table below are used to show the shortened versions of the commands if available. SA-2001

Page 15

14 Revision # Revisions Made Rev 01 Origination P:\MANUALS\BCGroup\…\SA2001_Series_UM_Rev01.docx

Page 16

15 MAINS VOLTAGE RANGE 90 to 264 VAC, 50/60 Hz ACCURACY ± 3% of Reading, ± 1 digit DEVICE CURRENT RANGE 0 to 19.99 Amps, RMS ACCURACY ± 5% o

Page 17 - NOTICE – PERFORMING TESTS

16 ELECTRICAL OPERATING LINE VOLTAGE 90 to 264 VAC, 50/60 Hz LINE PLUG SA-2001 MODELS NEMA 5-15P SA-2001-INTL MODELS Universal IEC C20 Receptacle Pig

Page 18

17 ELECTRICAL (continued) DUT RECEPTACLE (continued) SA-2001-INTL MODELS o EU1-16P (EURO) CEE 7/7 “SCHUKO” (NOTE: MUST USE BC20-20221

Page 19

18 NOTES

Page 21

BC GROUP INTERNATIONAL, INC. 3081 ELM POINT INDUSTRIAL DRIVE ST. CHARLES, MO 63301 USA 1-800-242-8428 1-314-638-3800 www.bcgroupi

Page 22 - MANUAL REVISIONS

i WARNINGS, CAUTIONS, NOTICES ... ii DESCRIPTION ...

Page 23 - SPECIFICATIONS

ii WARNING - USE The SA-2001 is intended for testing only and should never be used in diagnostics, trea

Page 24 - ELECTRICAL

iii CAUTION - SERVICE The SA-2001 is intended to be serviced only by authorized service personnel. Troubleshooting and service procedures should

Page 25 - PHYSICAL & ENVIRONMENTAL

iv NOTICE – SYMBOLS Symbol Description Caution (Consult Manual for Further Information) Electrical Caution (Consult Manual for Further

Page 26

v SA-2001 Series User Manual Copyright © 2012 www.bcgroupintl.com Made in the USA 08/12

Page 27

vi This Page Intentionally Left Blank

Page 28

1 The Model SA-2001 Series is a Microprocessor based Electrical Safety Analyzer. It allows for a multitude of tests to be performed on a de