TSI Certifier FA
TSI Certifier FA
Certifier FA Gas Range Accuracy**
Flow—High Flow Module Air, O
2
0 to 300 slpm* ±2% or ±0.075 slpm
Air/O
2
Mixtures 0 to 300 slpm ±4%or±0.1slpm
Flow—Low Flow Module Air, O
2
0.01 to 15 slpm ±2% or ±0.01 slpm
N
2
O 0.01 to 15 slpm ±4% or ±0.025 slpm
Volume-High Flow-Inhaled Air, O
2
0.01 to 10 liters STP ±2% Plus 0.02 liters
Air/O
2
Mixtures 0.01 to 10 liters STP ±4% Plus 0.02 liters
Volume-Low Flow-Inhaled Air, O
2
0.01 to 9.999 liters STP ±2% or ±0.01 liters
N
2
O 0.01 to 10 liters STP ±4% or ±0.01 liters
M
inute Volume-High Flow
0
.01 to 99 liters STP
±
7%
Minute Volume-Low Flow 0 to 9.999 liters STP ±7%
Inspiratory Time 0.25 to 60 seconds ±0.01 seconds
I:E Ratio– High Flow 1:100 to 100:1 ±5%
I:E Ratio– Low Flow 1:15 to 15:1 ±5%
Respiratory Rate 0.5 to 120 breaths per minute ±5%
Low Pressure -25 to +150 cmH
2
O ±0.75% or ±0.2 cmH
2
O
Barometric Pressure 7 to 29 PSI (500 to 2000 mbar) ±0.16 PSI (11 mbar)
Oxygen Concentration 21% to 100% 2% of concentration
To Order Certifier FA
Model Description
4070 High-Flow Standard Kit
Optional Modules and Accessories
Model Description
4073 Oxygen Sensor Kit
PSR-11-917-J Replacement Oxygen Sensor
BC20-40701 Replacement Low Flow Filter
BC20-40702 Replacement High Flow Filter
4072 Low-Flow Module
Part # Description
1319288 Hard shell carrying case
DPM-235175NNFC Digital Pressure Meter
Specifications
56
56
Flip for product pricing - Page U
Flip for product pricing - Page U
*slpm = Standard Liters per Minute **Accuracy stated as a percent of reading at TSI standard gas conditions.
Commentaires sur ces manuels