SYSTem Subsystem
This group allows the user to setup the startup mode for the unit, as well as directly
control the unit, as if pressing the keys on the front panel.
KEYWORD PARAMETER FORM COMMENTS
SYSTem:
POWer DEFaults | CUStom | SETCurrent
CONtrast < numeric_value > Numbers 1-20
KEY DUP | DDN | SUP | SDN | VUP |
VDN | SETup | RETurn | GSAVe |
HOLD
VER? Read only
READ Subsystem
This group allows the user to get measurements from the unit.
KEYWORD PARAMETER FORM COMMENTS
READ:
MVrms | MArms | Warms | MVPeak |
MVPP | MVP-PP | MVPK+ | CF |
TON | TOFF | TCYC | DCYC |
MVCyc | MACyc | WCyc Read only
DATA Read only
LOCn Read only, ‘n’ = stored
waveform location to read (1-3)
STATus Sub-system
This subsystem provides status on the operating mode of the unit including messages
that would normally be seen on the display.
KEYWORD PARAMETER FORM COMMENTS
STATus? Read Only
Bit
Value
Definition
0
1
Hold Mode
1
2
Graph Mode
2
4
Calibration Mode
3
8
4
16
5
32
6
64
7
128
8
256
Error Present
9
512
mV Out of Range
10
1024
mA Out of Range
11
2048
Watts Out of Range
12
4096
mV Peak Out of Range
13
8192
Crest Factor Out of Range
14
16384
15
32768