4-5
STEP 3
9
Choose Model of Unit Being Tested
Use to go to MODEL VIEW screen to enter
Manufacturer, Model, Serial Number, Department and
Location. (See MODEL SELECTION for details.)
CHANNEL 1 MODEL -VIEW
MANUF: ABC
MODEL: 123A
S/N: 12345
Choose Test
Scroll to DISTAL OCCLUSION test using .
CHANNEL 1 IDLE
SELECT TEST TYPE
DISTAL OCCLUSION
STEP 4
9
GETTING READY FOR A DISTAL OCCLUSION TEST
INFUSION INFUSION
INFUSION INFUSION
INFUSION
ANAN
ANAN
AN
ALAL
ALAL
AL
YZER--OCCLYZER--OCCL
YZER--OCCLYZER--OCCL
YZER--OCCL
USIONUSION
USIONUSION
USION
From the MODEL VIEW screen, use to return to the
test.
NOTES:
C
Each test is identified by time, date
and record number.
C
A test’s identification can be further
enhanced by entering model and
location information to be included in
the Datalogger and printouts.
C
This step is optional and can be
bypassed by simply moving on to the
next step.