BC Biomedical IPA-2000S Manuel d'utilisateur Page 60

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 162
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 59
2-28
PURGING OVERVIEW
PURGING OVERVIEW
PURGINGPURGING
PURGINGPURGING
PURGING
This section covers how to Purge the unit after use. Only
Distilled Water is to be used in the system; this should keep it
clean. However, if no fluid tests are going to be run for a
period of more than 24 hours, the water must be blown out of
the unit to maintain the necessary cleanliness.
PURGING
CAttach Clean Dry Air (CDA) or N
2
Line to Fluid In on Module
CAllow CDA or N
2
to Flow at 20 PSI for At Least One Minute
CMove Line to Fluid Out on Module
CGo Into User Setup Mode
CSet Air Purge Parameter to On
CWait at Least 2 Minutes Before Removing Line
The following is an outline of the steps that are involved in
purging the unit:
Vue de la page 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 161 162

Commentaires sur ces manuels

Pas de commentaire