RUNNING A PROXIMAL OCCLUSION TEST
Begin Test
Occlude the upstream fluid path.
Press to begin the timer.
STEP 1
9
STEP 2
9
Stop Test
Press to stop the timer. This should be done when
the pump alarm sounds. Insure the pump stops and/or
performs such action as called out by the pump manufacturer.
RUNNING A DISTAL OCCLUSION TEST
CHANNEL 1 PROX OCCL
ELAPSED TIME 00:00:00
CHANNEL 1 PROX OCCL
ELAPSED TIME XX:XX:XX