2-24
The following is an outline of the steps that are involved in
priming the unit:
PRIMING
CBegin from a Prime Channel Screen
CAttach an External Inlet Valve
CAttach Priming Syringe to the Valve
COpen Valve and Forcefully Inject 50 ml of Distilled Water
CClose Valve and Remove Priming Syringe
CPress Start Key to Set the Prime
CClear air from Device Under Test (DUT) Tubing
CConnect DUT Tubing Insuring there are No Bubbles
CRun 10 mL of Distilled Water through DUT and Module
PRIMING OVERVIEW
PRIMINGPRIMING
PRIMINGPRIMING
PRIMING