6-7
SAFETY ANALYZER AUTO
LINE VOLTAGE CHECK
Line Voltage
To ensure normal operating conditions for the DUT, the
Line Voltage is measured in this step. The display will
indicate if Line Voltage Passed.
(See ALARMS for details if it does not pass).
NOTE:
CThe Line Voltage must be within
+20% to –10% of its rated value.
STEP 4
9
STEP 5
9
SAFETY SAFETY
SAFETY SAFETY
SAFETY
ANAN
ANAN
AN
ALAL
ALAL
AL
YZER--AYZER--A
YZER--AYZER--A
YZER--A
UTUT
UTUT
UT
OMAOMA
OMAOMA
OMA
TIC MODETIC MODE
TIC MODETIC MODE
TIC MODE
RUNNING AN AUTOMATIC TEST
SAFETY ANALYZER AUTO
GND RESIST INVALID
LIMIT 2.00
MEASUREMENT --.--
Ground Lead Verification
To help confirm the ground lead is properly zeroed and
connected, the ground resistance measurement is
verified in this step. The display will indicate if Ground
Resistance is invalid and will pause the test until the
needed correction is made.
NOTES:
CGround Resistance would be invalid
if:
-the lead is not properly attached to
the DUT
-the lead is still attached to stud
-the test lead is bad or broken
C If the lead is adjusted and the
measurement becomes valid, the test
will continue.